Siemens LU01 User Manual

Page of 114
Page 82
SITRANS LU01 – INSTRUCTION MANUAL
7ML19985BE02
Pa
ra
m
e
te
rs
P907 Programmer Interface 
Press ENTER 
 
to activate the programmer interface (two way infrared 
communications) test. 
On successful test completion, PASS is displayed. Otherwise, FAIL is displayed. 
P910 Relay
Access this parameter to display the current state of the specified relay.
Press ENTER 
to alter the relay state (energized / de-energized) or enter the value 
corresponding to the specific relay state desired.
P911 mA Output Value 
Access this parameter to display the current value of the mA output.
Additionally, this feature may be used to enter a desired value. The mA output 
immediately assumes the value entered regardless of any restrictions programmed.
P912 Transducer Temperature
Use this feature to display the temperature in °C (as monitored by the connected 
transducer). 
Err is displayed if the transducer is not equipped with an internal temperature sensor.
P913 Sensor Temperature
Access this parameter to display the temperature in °C (as monitored by the TS-3). 
OPEn is displayed if a TS-3 is not connected.
Values
0 = de-energized
1 = energized
Primary Index
mA output
Values
Range: 0.00 to 25.00
Related
P200 mA Output Range
P201 mA Output Function
Values
Range: -50 to 150
Values
Range: -50 to 150