Agilent Technologies 66309B User Manual

Page of 190
 
Language Dictionary - 8
123
 
STATus:OPERation:NTR
STATus:OPERation:PTR
 
These commands set or read the value of the Operation NTR (Negative-Transition) and PTR (Positive-
Transition) registers. These registers serve as polarity filters between the Operation Enable and
Operation Event registers to cause the following actions:
u When a bit in the Operation NTR register is set to 1, then a 1-to-0 transition of the corresponding bit
in the Operation Condition register causes that bit in the Operation Event register to be set.
u When a bit of the Operation PTR register is set to 1, then a 0-to-1 transition of the corresponding bit
in the Operation Condition register causes that bit in the Operation Event register to be set.
u If the same bits in both NTR and PTR registers are set to 1, then any transition of that bit at the
Operation Condition register sets the corresponding bit in the Operation Event register.
u If the same bits in both NTR and PTR registers are set to 0, then no transition of that bit at the
Operation Condition register can set the corresponding bit in the Operation Event register.
 
Command Syntax
 
STATus:OPERtion:NTRansition<NRf>
STATus:OPERtion:PTRansition<NRf>
 
Parameters
 
0 to 32767
 
Preset  Value
 
NTR register = 0; PTR register = 32767
 
Examples
 
STAT:OPER:NTR 32         STAT:OPER:PTR 1312
 
Query Syntax
 
STAT:OPER:NTR?           STAT:OPER:PTR?
 
Returned Parameters
 
<NR1> (register value)
 
Related Commands
 
STAT:OPER:ENAB
 Table 8-5. Bit Configuration of Questionable Status Registers
 
Bit Position
 
15
 
14
 
13
 
12
 
11
 
10
 
9
 
8
 
7-6
 
5
 
4
 
3
 
2
 
1
 
0
 
Bit Name
 
not
used
 
Meas
Ovld
 
not
used
 
OC2
 
not
used
 
UNR
 
RI
 
UNR
2
 
not
used
 
SD
 
OT
 
FP
 
not
used
 
OCP
 
OV
 
Bit Weight
 
 
16384
 
 
4096
 
 
1024
 
512
 
256
 
 
32
 
16
 
8
 
 
2
 
1
 
OV = overvoltage protection has tripped on the main output (output 1)
 
OCP = overcurrent protection has tripped on the main output (output 2)
 
FP = the front panel "Local" key has been depressed
 
OT = overtemperature protection has tripped
 
SD = opened sense lead detected
 
UNR2 = output 2 is unregulated (Agilent 66309B/D only)
 
RI = remote inhibit is active
 
UNR = the output is unregulated
 
OC2 = output 2 overcurrent protection has tripped (Agilent 66309B/D only)
 
Meas Ovld = measurement overload
 
STATus:QUEStionable?
 
This query returns the value of the Questionable Event register. The Event register is a read-only register
which holds (latches) all events that are passed by the Questionable NTR and/or PTR filter.  Reading the
Questionable Event register clears it.
 
Query Syntax
 
STATus:QUEStionable[:EVENt]?
 
Parameters
 
None
 
Examples
 
STAT:QUES?
 
Returned Parameters
 
<NR1> (register value)
 
Related Commands
 
*CLS   STAT:QUES:ENAB  STAT:QUES:NTR   STAT:QUES:PTR