Siemens Software version syngo CT 2005C User Manual
192
Children
HeadRoutine_Baby
Indications:
Routine head spiral studies, e.g., brain tumors, cranial
trauma, hydrocephalus, hemorrhaging and other
abnormalities etc.
trauma, hydrocephalus, hemorrhaging and other
abnormalities etc.
A range of 8 cm will be covered in 27 sec.
* The conversion factor for an 8-week-old child, and a
scan range of 80 mm was used.
scan range of 80 mm was used.
Head
kV
80
Effective mAs
122
Rotation time
1.5 sec
Slice collimation
2.5 mm
Slice width
3.0 mm
Feed/Rotation
5.0 mm
Pitch Factor
1.0
Increment
3.0 mm
Kernel
C30s
CTDI
Vol
7.89 mGy
Effective dose
Male: 0.84 mSv*
Female: 0.92 mSv*
Contrast medium IV injection
Start delay
Start delay
exam dependent
Flow rate
dependent upon needle size/
Access site
Total amount
1 – 2 ml per kg of body weight
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 192 Friday, April 8, 2005 9:55 AM