Children
ExtrHR_Child
Indications:
Spiral mode for bone studies, e.g., tumors, post-
traumatic changes, orthopedic indications, etc.
A range of 10 cm will be covered in 27 sec.
* The conversion factor for a 7-year-old child, and a 
scan range of the knee of 100 mm was used.
ExtrHR
kV
130
Effective mAs
40
Rotation time
1.0 mm
Slice collimation
1.0 mm
Slice width
1.25 mm
Feed/Rotation
4.0 mm
Pitch Factor
2.0
Increment
0.8 mm
Kernel
B70s
CTDI
Vol
4.31 mGy
Effective dose
Male: < 0.01 mSv*
 
Female: < 0.01 mSv*
Contrast medium IV injection
Start delay 
exam dependent
Flow rate 
dependent upon needle size/
Access site
Total amount 
1 – 2 ml per kg of body weight