Siemens Software version syngo CT 2005C User Manual

Page of 321
230
 
Children
CarotidAngio
Indications:
Spiral mode for the carotid arteries, e.g., carotid steno-
sis or occlusion, vascular abnormalities of the carotids 
or vertebral arteries, etc.
A range of 10 cm will be covered in 27 sec.
* The conversion factor for a 7-year-old child, and a 
scan range of 100 mm was used.
 CarotidAngio
2
nd
 reconstr.
kV
80
Effective mAs
50
Rotation time
1.0 sec
Slice 
collimation
1.0 mm
Slice width
3.0 mm
1.25 mm
Feed/Rotation
4.0 mm
Pitch Factor
2.0
Increment
3.0 mm
0.8 mm
Kernel
B41s
B41s
CTDI
Vol
1.38 mGy
Effective dose
Male: 
0.50 mSv*
 
Female: 
0.50 mSv*
Contrast medium IV injection
Start delay 
exam dependent
Flow rate 
dependent upon needle size/
Access site
Total amount 
1 – 2 ml per kg of body weight
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book  Page 230  Friday, April 8, 2005  9:55 AM