Head
71
 
Hints in General
• Topogram: Lateral, 256 mm.
• Patient positioning:
 
Patient lying in supine position, arms resting against 
body, secure head well in the head holder, support 
lower legs.
• Gantry tilt is available for both, sequence and spiral 
scanning.
 
However, image artifacts may occur if spirals are 
acquired with a tilt angle greater than 8°.
• For all head studies, it is very important for image 
quality to position the patient in the center of the 
scan field. Use the lateral laser beam to make sure 
that the patient is positioned in the center.
• In order to optimize image quality versus radiation 
dose, scans are provided within a maximum scan 
field of 300 mm with respect to the iso-center. No 
recon job with a field of view exceeding those limits 
will be possible. Therefore, patient positioning has to 
be performed accurately to ensure a centered loca-
tion of the skull.
Head Kernels
• For soft tissue head studies, the standard kernel is 
H41s; softer images are obtained with H31s or H21s 
sharper images with H50s. 
• High resolution head studies should be performed 
with H60s, H70s (e.g., for dental and sinuses) and 
H80s (e.g., inner ear).