Siemens Software version syngo CT 2005C User Manual

Page of 321
78
 
Head
InnerEarHRSeq
Indications:
Sequential mode for inner ear studies, e.g., inflamma-
tory changes, tumorous processes of pyramids, cere-
bellopontine angle tumors, post-traumatic changes, 
etc.
A typical gantry tilt of -20 degree is predefined. A 
range of 2.5 cm will be covered.
InnerEarSeq
kV
130
Effective mAs
140
Rotation time
1.5 sec
Slice collimation
1.0 mm
Slice width
1.0 mm
Feed/Scan
2.0 mm
Kernel
H80s
CTDI
Vol
31.88 mGy
Effective dose
Male: 0.32 mSv
 
Female: 0.38 mSv
Contrast medium IV injection
Start delay
60 sec.
Flow rate
2 ml/sec.
Total amount
50 – 60 ml