Agilent Technologies 81480A User Manual

Page of 194
Test Record
Performance Tests
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide, Sixth Edition
169
Agilent 81689A Performance Test
         Page 5 of 6
Model Agilent 81689A Tunable Laser
Report No. ________ Date_______
Maximum Power Test
Measurement Uncertainty: 
±0.10 dB 
Power Linearity
Maximum Power Linearity at current setting
______dB
Minimum Power Linearity at current setting
______dB
Total Power Linearity = (Max Power Linearity 
− Min Power Linearity)
______dBpp
Specification
    0.5 dBpp
typical
    0.2    dBpp
    
Measurement Uncertainty 
 
±0.05  dB
Wavelength Setting
Power Measured
Minimum Specification
1525.000 nm
dBm
+ 6.00 dBm
1535.000 nm
dBm
+ 6.00 dBm
1540.000 nm
dBm
+ 6.00 dBm
1550.000 nm
dBm
+ 6.00 dBm
1560.000 nm
dBm
+ 6.00 dBm
1570.000 nm
dBm
+ 6.00 dBm
1575.000 nm
dBm
+ 6.00 dBm
Power Setting from 
start
Measured Relative  
Power from start
Power reduction 
from  start
Power Linearity at cur-
rent setting
Start = REF
+ 6.00 dBm
    0.00 dB 
+
0.00 dB
=
    0.00 dB
+ 5.00 dBm
dB +
1.00 dB
=
dB
+ 4.00 dBm
dB +
2.00 dB
=
dB
+ 3.00 dBm
dB +
3.00 dB
=
dB
+ 2.00 dBm
dB +
4.00 dB
=
dB
+ 1.00 dBm
dB +
5.00 dB
=
dB
−  0.0 dBm
dB +
6.00 dB
=
dB
−  1.0 dBm
dB +
7.00 dB
=
dB
−  2.0 dBm
dB +
8.00 dB
=
dB
−  3.0 dBm
dB +
9.00 dB
=
dB