Agilent Technologies 81480A User Manual
Performance Tests
Test Record
170
Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules User’s Guide,
Sixth Edition
Agilent 81689A Performance Test
Page 6 of 6
Model Agilent 81689A Tunable Laser
Report No. ________ Date_______
Power Flatness
Power Stability
1
Power Stability = Maximum Deviation – Minimum Deviation
Signal-to-Source Spontaneous Emission
Measurement Uncertainty:
±0.20 dB
Wavelength
P = +2.0 dBm
Power Deviation
Start = REF
1525 nm
0.00 dB
1530 nm
dB
1540 nm
dB
1550 nm
dB
1560 nm
dB
1570 nm
dB
1575 nm
dB
Maximum deviation
dB
Minimum deviation
dB
Flatness =
Maximum
− Minimum Deviation
dB
Specification 0.60
dBpp
Measurement Uncertainty
± 0.10 dB
P =
−3.0 dBm
Maximum Deviation
dB
Minimum Deviation
dB
Power Stability
1
dB
Specification
0.06 dBpp
Measurement Uncertainty
± 0.005 dB
Wavelength
Output Power
Results
Maximum
Specification
Typical
1525 nm
+6.00 dBm
dB
30 dB
39 dB
1535 nm
+6.00 dBm
dB
30 dB
39 dB
1545 nm
+6.00 dBm
dB
30 dB
39 dB
1555 nm
+6.00 dBm
dB
30 dB
39 dB
1565 nm
+6.00 dBm
dB
30 dB
39 dB
1575 nm
+6.00 dBm
dB
30 dB
39 dB