Sierra 840 User Manual

Page of 74
Chapter 4 Maintenance 
Series 830/840/860 Instruction Manual
 
4-2 
IM-83/84/86-H
 
 
Figure 4-1. Low Flow Transducer LFE Cleaning 
Low Flow Transducers: 
The LFE is accessed by unscrewing the main inlet fitting and re-
moving it from the flow body.  The LFE is screwed into the inlet 
fitting, which has been specially machined for this purpose. To ac-
cess the components: 
 
1.  Remove the unit from the system. 
 
2.  The inlet filter screen is held in place in the inlet fitting by the 
LFE. Disassemble by holding the fitting steady with a wrench 
and unscrewing the LFE with a medium flat-tipped screw-
driver. 
 
3.  Remove the LFE assembly taking care not to bend the inlet 
screen. Inspect the sealing O-ring and replace if necessary. In-
spect the inlet screen and replace if corroded or damaged. Light 
to medium particulate contamination can be cleaned by back 
washing with a suitable solvent. Air dry thoroughly. 
 
4.  Inspect the LFE for damage and replace if necessary. Replace-
ment of the LFE or inlet screen requires transducer re-
calibration. 
 
5.  Re-assemble components. When the transducer is re-installed 
in the system, leak test the connection. 
 
 
6.  Check transducer calibration.  
 
Caution! 
Removing or re-
positioning the LFE may 
alter the calibration..