Sierra 840 User Manual

Page of 74
Series 830/840/860 Instruction Manual 
Chapter 4 Maintenance
 
IM-83/84/86-H 
4-3
 
 
Figure 4-2. Medium Flow Transducer LFE Cleaning 
Medium Flow Transducers: 
In the medium flow body, the LFE assembly consists of the honey-
comb laminar flow element, inlet screen, 0.63 inch long standoff, 
two ranging washers, 2-1/4 inch long 4-40 screw and 4-40 nut. 
Range changes in the honeycomb element are made with various 
diameter ranging washers. To access the components: 
 
7.  Remove the unit from the system. 
 
8.  Access the LFE by unscrewing the four 10-32 socket head cap 
screws from the inlet side of the flow body and remove the in-
let end cap. 
 
9.  Remove the LFE assembly taking care not to bend the inlet 
screen. Inspect the sealing O-ring and replace if necessary. In-
spect the inlet screen and replace if corroded or damaged. Light 
to medium particulate contamination can be cleaned by back 
washing with a suitable solvent. Air dry thoroughly. 
 
10. Inspect the honeycomb element for damage and replace if nec-
essary. Replacement of the LFE or inlet screen requires trans-
ducer re-calibration. 
 
11. Re-assemble components. When the transducer is re-installed 
in the system, leak test the connection. 
 
12. Check transducer calibration.