Sierra 840 User Manual

Page of 74
Chapter 5 Troubleshooting 
Series 830/840/860 Instruction Manual
 
5-2 
IM-83/84/86-H
 
Flow Meters and Flow Controllers with Electromagnetic Valves 
 
 
Problem 
Possible Cause 
Solution 
 
 
 
No output 
Clogged sensor 
 
PCB defective 
 
Inlet filter screen clogged 
Clean or replace sensor 
 
Repair or replace PCB 
 
Clean or replace screen 
 
 
 
Unit will not zero 
Gas leak 
 
Application requires high pressure and 
non-horizontal mounting 
 
PCB defective 
Find and correct leaks 
 
Re-zero transducer 
 
 
Repair or replace PCB 
 
 
 
Controller does not 
respond to set point 
Low or no gas pressure 
 
Faulty cable or connector 
 
Set point is below 2% of full scale 
Set correct gas pressure 
 
Correct or replace 
 
Increase set point or disable auto 
shut off circuit 
 
 
 
Flow does not match set point 
No gas pressure 
 
Inlet filter screen clogged 
 
Out of adjustment 
Set correct gas pressure 
 
Clean or replace 
 
Adjust R22 balance on 840 models, 
consult factory for others 
 
 
 
Reads full scale with no flow 
or with valve shut 
Defective sensor 
 
Gas leak 
Return to factory for replacement 
 
Find and correct leaks 
 
 
 
Out of calibration 
Dirty or clogged sensor 
 
Change in composition of gas 
 
Gas leak 
 
PCB defective 
 
LFE dirty 
 
Inlet filter screen clogged 
 
Incorrect inlet conditions (high flow and 
NPT models) 
Clean or replace sensor 
 
See K-factory tables 
 
Find and correct leaks 
 
Repair or replace PCB 
 
Clean LFE 
 
Clean or replace screen 
 
Re-plumb transducer correctly