Counting Mode
8 Counting Mode
In counting mode the total weight of articles is converted to a count and assumes
all articles have the same mass value.
It is necessary to store a unit mass to count articles.
8.1  Storing a Unit Mass
Step   1   Press the  
MODE
  switch to display the unit  
PCS
.
Step 2 Press the 
 SET 
 switch to enter the mode to store a
unit mass.
Step 3 Press the 
 
^
 
 switch to select the number of sam-
ples. The greater the quantity of samples, the
greater the accuracy of the count.
5 pieces, 10 pieces, 20 pieces, 50 pieces, 100
pieces
Step 4 Put the container onto the pan.
When the stability mark is displayed press the
TARE 
 switch.
Step 5 Put in the number of samples selected at step 3.
      
Press the  
ENTER
  switch to store the weight after
      the stability mark is displayed. The count is then
       displayed.
Caution
When the sample is too light and it is not
 
possible to
calculate a unit mass, the scale displays 
 lo ut 
 and
returns to step 3. It is necessary to increase the
number of samples.
Pressing the  
ENTER
  switch after the 
 lo ut .
 is
displayed, the next weighing mode is displayed.
When the unit mass is too light to store, the scale
displays 
 lo ut .
Step 6
Remove all items from the pan.