A&D HV-G Series User Manual

Page of 69
HV-G Instruction Manual Page 23
9 Percentage Mode
The percent mode is to display a mass value in the unit of “%”.
It is necessary to store a 100% mass value, in advance to use this function.
9.1 Storing a 100% Mass
Step 1
Press the  
MODE
  switch to display the unit %.
Step 2
Press the  
SET
  switch to enter the mode that
stores a 100% mass.
Step 3
With nothing on the pan press the  
ZERO
  switch
to display zero.
Step 4
Put the 100% mass on the pan gently.
Press the  
ENTER
  switch to store it after the
stability mark is displayed. Then the
percentage unit is displayed.
Caution
When the sample is too light and it is not possible
to calculate a 100% mass, the scale displays 
 lo .
and returns to step 3.
Pressing the  
ENTER
  switch after 
 lo 
 is displayed,
the next weighing mode is displayed.
Step 5
Remove all items from the pan.