A&D HV-G Series User Manual

Page of 69
HV-G Instruction Manual Page 38
Step 6
Press the  MODE  switch to display the blinking
HI (of the final value).
Step 7 Set the final value using the following switches.
 
^
 
 switch
Selecting the value of a digit.
 
<
 
 switch
Selecting a digit.
Step 8 Press the 
 ENTER 
 switch to store the new
parameter and display the blinking OK
(of preliminary value).
Step 9 Set the preliminary value using the following
switches.
 
^
 
 switch
Selecting the value of a digit.
 
<
 
 switch
Selecting a digit.
Step 10 Press the 
 ENTER 
 switch to store the new
parameter and display the blinking LO (of zero
band).
Step 11 Set the zero band using the following switches.
 
^
 
 switch
Selecting the value of a digit
 
<
 
 switch
Selecting a digit.
Step 12
Press the  
ENTER
  switch to store the new
parameter and display the weighing mode.
13.2 Operation and Performance (Examples)
Step 1 Select the parameter 
 f6 8 
 of the function table.
Step 2 Set the parameters of the final value, preliminary value and zero band.
Step   3  The comparison result is always displayed.