17 Options
17.1 RS-232C / Relay output / Buzzer (OP-03)
Replacing RS-232C interface with this option, refer to “ RS-232C Serial Interface” for its
specification.
The following option cables can be used, when you do not use the relay output.
AX-KO557A,
AX-KO1786-200
Pin connections
Mating connector
DIN 8pin, JA:TCS0586 (of accessory pack)
Circuit
Relay
Solid-state-relay
Maximum voltage
DC50V
Maximum current
DC100mA
Maximum ON resistance 8 
Ω
Pin No. Signal name
Direction
Description
1
HI
Output
Relay output of HI
2
RXD
Input
Receive data
3
TXD
Output
Transmit data
4
LO
Output
Relay output of LO
5
SG
-
Signal ground (RS-232C)
6
OK
Output
Relay output of OK
7
DSR
Output
Data set ready
8
COM
-
Relay common terminal
Installing OP-03
Step 1
Remove two screws that
attach the RS232C board on
the rear of the display unit.
Remove the RS232C board.
Step 2
Switch the connections.
Step 3
Insert the option board into the
display unit and afix with
screws