RS232 Serial Interface
Chapter 10    Peripherals    
111
NIBP Monitor
One of the following devices can be connected to an M1351A or M1353A to record 
maternal NIBP:    
l
Dinamap
1846/8100 NIBP Monitor. This requires a Critikon model 8801 adapter
l
COLIN Press-Mate/Nippon Colin Listmini Model BP-8800.
Before connecting a NIBP monitor, ensure that the correct configuration options on the 
M1351A and M1353A have been selected. See Table 4-2 on page 27 for the options 
available. Interface to an NIBP Monitor is available only on the Combined Interface 
Module, Option J13, with the label M1353-66531, E (software revision number A.02.xx or 
higher - the software revision number is written on the error log).
                      
Connecting
an NIBP Monitor to
the M1351A or
M1353A
To connect a Dinamap
 1846/8100 NIBP Monitor to the M1351A or M1353A:
1. Use the interface cable M1350-61609.
2. Connect the 9-pin end of the cable to the fetal monitor.
3. Connect the 25-pin end of the cable to the external device.
4. Secure the interface cable by the screws at the cable connectors.
Figure 10-5Interface Cable M1350-61609
Caution
Before connecting an NIBP Monitor, use the Grounding Cable (part number 
8120-2961 for USA or 8120-4808 for Europe) to connect the equipotential 
grounding point on the fetal monitor to earth potential. The power cord of 
the NIBP Monitor must be plugged into a wall-mounted power outlet - not 
into an extension block.