Philips M1351A User Manual

Page of 189
RS232 Serial Interface
Chapter 10    Peripherals    
113
FSpO
2
 Monitor
If the fetal monitor has the Combined Interface Module J-14 fitted, you can interface to an 
FSpO
2
 monitor (Nellcor N-400 or other compatible). You must have software revision 
number A.02.09 or higher. 
To connect an FSpO
2
 monitor to the M1351A or M1353A:
   
1. Use interface cable M1353-61614. This has identical 9-pin male connectors at each 
end.
2. Connect one end to the M1351A or M1353A and the other to the FSpO
2
 monitor.
3. Secure the interface cable by the screws at the cable connectors.
Figure 10-7FSpO
2
 Interface Cable (M1353-61614)
Baudrates
The baudrates for the FSpO
2
 monitor must be set as follows:
Baudrate:
2400
Startbit:
1                          Parity:                None
Databits:
8
Stopbits:
1
Cable:
M1353-61614
Table 10-3 NIBP Monitor and Minimum Time Interval
Paper Speed
Minimum Time Interval
1 cm/min
3 minutes.
2 cm/min
2 minutes.
3cm/min
1 minute.