eutech-instruments cyberscan ph 110 phmvrs232 meter User Manual

Page of 76
Instruction Manual   
 
CyberScan pH 11 / 110  
 
25
8.2.1  Print Current Data Manually 
To manually print any data that is currently being 
measured:  
1.  From measurement mode, press HOLD key to 
freeze reading. See Figure 24.  
2.  Print PRINT (
) key.  
3.  The “Prn – CUr” annunciator will appear and 
flash for about 0.5 seconds. The data is 
transmitted to the printer or computer.  
4.  The meter returns to measurement mode.  
5.  Repeat steps 1 to 3 for next measurement.  
 
8.2.2  Print Data on Timed Interval 
This feature is useful to automate printing and 
downloading ‘live’ measurement values into computer. 
The time interval setting can range from 5 to 30 
seconds.  
1.  From pH or mV measurement mode, press 
PRINT key.  
2. Select “Dat/Cur” screen and press 
HOLD/ENTER key.  
3.  The “Sec 5” screen appears. The default value 
is 5 seconds time interval. See Figure 25.  
4. Press 
MI/▲ or MR/▼ key to set a value suitable for your application.  
5.  Press HOLD/ENTER key.  
6.  The meter goes into measurement mode with MEM annunciator flashing.  
7.  At a pre-set time interval, the screen flashes “Prn/CUr” momentarily (printing 
data) and switches back to measurement mode with MEM annunciator 
flashing.  
8.  To stop printing, press PRINT key. The MEM annunciator disappears and the 
meter is on measurement mode.  
 
pH
pH
MEAS
MEAS
READY
READY
 23.4
 23.4
6.55
6.55
ENTER
HOLD
°C
°C
pH
MEAS
READY
HOLD
Flash for 0.5 seconds
 23.4
6.55
°C
Figure 24: Print current data 
manually