9.2.3  P1.3 Select Number of pH Calibration Points 
Program P1.3 lets you select the number of calibration points that appear in pH 
calibration mode: 2, 3, 4, or 5 (or 6 with DIN buffer option). This lets the meter scroll 
through the calibration points more quickly if you regularly calibrate at less than 5 
points.  
 From measurement mode: 
1.  Press SETUP key to enter Set Up mode. 
2.  Press the MI/▲ or MR/▼ keys to scroll through 
subgroups until you view parameter P1.0.   
3.  Press the HOLD/ENTER key three times to 
select parameter 1.3.  
4.  Press the MI/▲ or MR/▼ keys to select 2, 3, 4, 
or 5 point pH calibration (up to 6 points for DIN 
calibration set with pH 110 meter).  
5. Press the HOLD/ENTER key to confirm 
selection and to return to the subgroup menu.  
Press CAL/MEAS key to return to measurement 
mode.  
See Figure 32.  
 
 
 
 
Figure 32: P1.3 - Select 
number of calibration points