P2.0: Viewing Previous Calibration Data  
This mode lets you recall previous calibration data, 
which helps you know when to re-calibrate your meter.  
This is a “view only” mode.  
From measurement mode: 
1.  Press the SETUP key to enter Set up mode.  
2.  Press the MI/▲ or MR/▼ keys to scroll through 
subgroups until you view parameter P2.0. 
3.  Press the HOLD/ENTER key repeatedly to 
view previous calibration data.  See Figure 34. 
4.  When you have scrolled through all calibration 
data, you will automatically return to the 
subgroup menu. Press CAL/MEAS key to 
return to measurement mode.  
 
 
 
 
 
 
 
 
 
NOTES: 
If there is no previous calibration data at a particular point, the 
primary display will show “ ----“.  
 
 
 