NOTE:  
Calibration reset is independent for each measurement parameter. 
That is:  
•  From pH mode, calibration reset will only erase pH calibration 
values; and 
•  From mV mode, calibration reset will only erase mV calibration 
values.  
9.6.2 User 
Reset 
This program clears all data including calibration, 
memory and other customised setup functions and reset 
it to factory default.  
You can skip program P5.1 (Calibration Reset) and 
proceed straight to Program P5.2: 
1.  From Program P5.0 - rSt, press HOLD/ENTER 
key three times to P5.2.  
2. Press 
the 
MI/▲ or MR/▼ keys to scroll through 
subgroups until you view parameter P5.0.  
• 
‘NO’ retains current settings 
• 
‘YES’ resets to factory default settings.  
3.  Press the ENTER key to confirm selection.  
•  If ‘NO’ is selected, it will retains current 
settings and goes back to program 5.0.  
• 
If ‘YES’ is selected, all LCD segments will lit 
up for one second, resets to factory default 
settings, and the meter will be in the 
measurement mode.  
4.  Otherwise press CAL/MEAS key to return to 
measurement mode without resetting to factory 
default.  
 
 
 
yes
SETUP
SETUP
SETUP
 P 5.2
ENTER
HOLD
MI
MR
Select either "YES" or "nO", 
press ENTER key to confirm. 
Figure 39: P5.2 - User reset