eutech-instruments cyberscan ph 110 phmvrs232 meter User Manual

Page of 76
Instruction Manual   
 
CyberScan pH 11 / 110  
 
60
12 ERROR 
MESSAGES 
The following table provides a guideline to enable diagnosis of possible problems 
indicated by the messages generated by the meter. The table also provides possible 
solutions to the problems encountered. 
ERROR MESSAGE 
INDICATES 
POSSIBLE CAUSE 
CORRECTIVE 
ACTION 
Err. Annunciator 
Wrong keypad input 
Wrong input in 
selected mode.  
Release key. 
Select valid 
operations 
depending on 
mode.  
Secondary Display 
Toggles Continuously 
between Calibration 
Buffer Values, 
Electrode icon  blinks & 
Err annunciator lights 
up 
Calibration  
Incorrect buffer used 
or contaminated buffer 
solution 
Check if the right 
buffer was 
selected or use 
fresh buffer 
solution. 
Battery icon lights up 
Low Battery 
Battery power is low 
Replace batteries 
with fresh ones as 
soon as possible.