CyberScan pH 11 / 110  
 
61
13 TROUBLE-SHOOTING 
 
PROBLEM PROBABLE 
CAUSE 
REMEDIAL 
ACTION 
Nothing is displayed when 
the ON/OFF key is 
selected. 
1.  Batteries not in place. 
2.  Batteries (+ & - poles) not 
in correct polarity. 
3. Weak 
batteries. 
1. Insert 
batteries. 
2.  Re-insert batteries in the 
correct polarity. 
3.  Replace batteries or 
attach AC adapter. 
 
Unstable reading 
1.  Insufficient reference 
electrolyte in electrode. 
2. Broken 
electrode. 
3.  External “noises” or 
induction (e.g. electrical 
‘noise’ caused by a 
nearby motor). 
4. Dirty 
electrode. 
1.  Fill electrode with 
reference electrolyte. 
2. Replace 
electrode. 
3.  Move or switch off 
interfering device. 
4.   Clean the electrode. 
Rejuvenate if necessary.  
Recalibrate the meter 
with the cleaned probe. 
Slow responses. 
1.  Dirty/Oily electrode. 
1.  Clean electrode. 
Rejuvenate if necessary. 
Meter is not responding to 
key press. 
1.  HOLD mode in operation. 
2.  Internal program error. 
1.  Cancel HOLD mode. 
2.  Reset all internal 
programs by re-inserting 
battery.