10.6 Remote Zero Adjustment 
On-stream zero adjustment in a zero flow rate condition can be started with an external signal.   
To do this, set DI as a zero adjustment start signal. See 8.2.19, “Digital I/O” 
 
[Signal Input Timing] 
 
 
 
 
 
 
 
                                                                                                                       
   
 
 
*
 The start signal must be set to H level first, then it must go to L level after the passage of more 
than 10 seconds but not more than 20 seconds, as shown above.   
If the signal does not go to L level within this specified period, it will be ignored. 
 
10.7 Remote Selection of Fixed Value Output 
A user-specified 4-20 mA output and pulse output can be selected with a DI signal
Proceed as follows to use this function: 
 
Fixed-value setting 
 
 
„ 
Set the fixed-value for current output and for pulse output. See 8.2.25, “Fixed-Value 
Output.” Set the fixed-value output enable/disable status to “OFF.” 
If the pulse output is not used, fixed-value setting for pulse output is not needed. 
 
DI function setting 
 
 
„ 
Set DI to use as a fixed-value output control signal. See8.2.19, “Digital Input.” 
 
 
Control signal input conditions: 
Control signal input level 
4 –20 mA and pulse output 
L level 
Outputs the measured value. 
H level 
Outputs the fixed-value. 
10 seconds 
20 seconds 
H level 
L level 
Zero adjustment 
start signal 
 
Zero adjustment start