eutech-instruments ammonium epoxy User Manual

Page of 16
Instruction Manual 
 
Ammonium Electrode 
 
13
TROUBLESHOOTING HINTS 
 
Symptom 
Possible Causes 
Next Step 
Out of Range 
defective meter 
check meter with Reading   
  
shorting 
strap 
 
 
(see meter instruction manual)   
                                      
 electrode(s) 
unplug 
electrode(s) 
 
not plugged in 
and reset 
 properly   
 
 
reference electrode 
be sure reference  
 
not filled  
electrode is filled 
 
 
air bubbles on  
remove bubble by 
 membrane 
redipping 
electrode 
 
 
electrodes not 
put electrode(s) 
 
in solution  
in solution     
 
 
"Incorrect  
incorrect scaling 
plot millivolts on the linear 
Answer" 
of semi-log paper 
axis. On the log axis, be sure 
curve is good) 
 
concentration numbers within 
 
(but calibration 
 
each decade are increasing 
 
 
 with increasing concentration 
 
 
incorrect sign 
be sure to note  
 
 
sign of millivolt number correctly 
 
 
incorrect standards 
prepare fresh standards 
 
 
 
wrong units used 
apply correct conversion factor:  
  
10-
3
M = 18 ppm NH
4
+ = 
 
 
14 ppm as N 
 
 
sample carryover 
rinse electrodes  
 
 
 
 
thoroughly between samples 
 
 
Drift (reading 
samples and standards 
allow solutions 
slowly changing 
at different  
to come to room 
in one direction) 
temperatures 
temperature  
  
before 
measurement 
 
 
electrode exposed 
soak electrode  
 
to interference 
in ammonium standard 
 
 
incorrect reference 
use recommended 
 
filling solution 
filling solution