eutech-instruments ammonium epoxy User Manual

Page of 16
Ammonium Electrode 
 
Instruction Manual 
 
14
 
Low Slope or 
standards contaminated 
prepare fresh 
No Slope 
or incorrectly made 
standards 
 
 
ISA not used 
use recommended ISA 
 
 
standard used as ISA 
use ISA 
 
 
defective electrode 
check electrode 
  
operation 
 
 
electrode exposed 
soak electrode 
 
to interference 
in ammonium standard 
 
 
air bubble on membrane 
remove bubble by 
  
redipping 
probe 
 
 
Noisy or Unstable 
defective meter 
check meter with 
Readings (readings 
 
shorting strap 
continuously or 
 
randomly changing) 
air bubble on  
remove bubble by 
 membrane 
redipping 
electrode 
 
 
defective electrode 
replace electrode 
 
 
 
 
electrode exposed 
soak electrode  
 
to interference 
in ammonium standard 
 
 
meter or stirrer 
ground meter or  
 not 
grounded 
stirrer 
 
 
outer filling solution 
fill electrode 
 
level too low  
to level just below fill hole