eutech-instruments waterproof cyberscan con 400410 & tds 400 conductivitytds meter User Manual

Page of 56
Instruction Manual 
 
CON 400/410 
- 9 - 
4.3 
Temperature Calibration  
Your probe features a built-in temperature sensor. The 
temperature sensor is factory calibrated. Calibrate your 
sensor only if you suspect temperature errors that may 
have occurred over a long period of time or if you have a 
replacement probe.  
 
Temperature Calibration  
1. 
Make sure the cell is attached to the 6-pin connector. 
The ATC annunciator will appear at the right-hand 
side of the LCD*.  
2. 
Switch the meter on. Press the MODE key to select 
conductivity or TDS mode.  
3. Press 
the 
CAL/MEAS key to enter conductivity or 
TDS calibration mode. The CAL indicator appears 
above the primary display.  
4. 
While in conductivity or TDS calibration mode, press 
the MODE key to enter temperature calibration 
mode. The primary display shows the current 
temperature reading and the secondary display 
shows the factory default temperature value.  
5. 
Dip the cell into a solution of known temperature (i.e. 
A temperature bath). Allow time for the built-in 
temperature sensor to stabilise.  
6. 
Scroll with the MI/
S or MR/T keys to set the correct 
temperature value (i.e. the temperature of the 
temperature bath). You can adjust the reading in 
increments of 0.1 
°C.  
7. 
Once you have selected the correct temperature, 
press the ENTER key. 
* If the ATC indicator does not light up, see Program P4.3 
to switch it on. 
 
NOTES: 
• 
You can offset the temperature reading up to 
±5°C from 
the original reading (default reading).  
• 
To exit this program without confirming the temperature 
calibration value, DO NOT press ENTER. Press 
CAL/MEAS instead.  
• 
If the ATC indicator does not light, see Program P4.3, 
on page 32 to switch it on. 
• 
Since temperature readings affect the accuracy of the 
conductivity/TDS measurements, it is strongly 
recommended to carry out a conductivity/TDS 
calibration after a temperature calibration is done. 
Figure 10 - Temperature 
Calibration 
MEAS
1413
1413
ATC
°C
22.3
ATC
°C
22.3
MEAS
ATC
°C
22.0
CAL
CAL
ATC
°C
22.3
22.3
ATC
°C
22.0
22.3
CAL
CA
L
ME
AS
READY
MR
MI
MO
DE
µS
µS
1425
µS
EN
TER
RA
NG
E