eutech-instruments waterproof cyberscan con 400410 & tds 400 conductivitytds meter User Manual

Page of 56
Instruction Manual 
 
CON 400/410 
- 7 - 
4   C A L I B R AT I O N 
 
4.1 
Important Information on Meter Calibration  
Your meter has five measuring ranges. You can calibrate one point in each of the measuring 
ranges (up to five points). If you are measuring values in more than one range, make sure to 
calibrate each of the ranges you are measuring.  
 
To view current calibration points, see SETUP section Program 2.0 on page 28.  
 
The following table lists the corresponding conductivity and TDS ranges. You should calibrate 
each range using a solution that falls between the values in the “recommended calibration 
solution range” column.  
 
Conductivity 
Measuring 
Range 
Recommended 
Calibration Solution 
Range 
TDS Measuring 
Range 
Recommended 
Calibration Solution 
Range 
0.00 
Æ 19.99 
µS 
6.00 to 17.00 
µS 0.00 
Æ 9.99 ppm 
3.00 to 8.50 ppm 
0.0 
Æ 199.9 
µS 
60.0 to 170.0 
µS 10.0 
Æ 99.9 ppm 
30.0 to 85.0 ppm 
Æ 1999 
µS 600 
to 
1700 
µS 100 
Æ 999 ppm 
300 to 850 ppm 
0.00 
Æ 19.99 mS 
6.00 to 17.00 mS 
1.00 
Æ 9.99 ppt 
3.00 to 8.50 ppt 
0.0 
Æ 199.9 mS 
60.0 to 170.0 mS 
10.0 
Æ 200 ppt 
30.0 to 170 ppt 
 
When you recalibrate your meter, old calibrations are replaced on a range by range basis. For 
example, if you previously calibrated your conductivity meter at 1413 
µS in the 0 to 1999 µS 
range and you recalibrate at 1500 
µS (also in the 0 to 1999 µS range), the meter will replace 
the old calibration data (1413 
µS) in that range. The meter will retain all calibration data in 
other ranges.  
 
To completely recalibrate your meter, or when you use a replacement probe, it is best to clear 
all calibration data in memory. To erase all the old conductivity and TDS calibration data 
completely from memory, see SETUP section Program 8.0 on page 39.  
 
For information on how to calibrate your meter: 
•  See section 4.3 for Temperature Calibration.   
•  See section 4.4 for Conductivity  
•  See section 4.5 for TDS Calibration (CON 410 meter only).   
•  See Addendum 1 for more calibration tips.