7.7 
P6.0 Selecting the cell constant 
Your meter lets you select a cell constant of K = 1.0, 10, or 0.1.  
Use a cell of K = 1.0 for midrange measurements 
Use a cell of K = 10 for high range measurements (above 20 mS or 10 ppt).  
Use a cell of K = 0.1 for low range measurements (below 20 
µS or 10 ppm).  
The cell included with your meter has a cell constant of K = 1.0.  
 
From measurement mode 
1. Press 
SETUP key to enter Set Up mode.  
2. Press 
the 
MI/
S or MR/T keys to scroll through 
subgroups until you view parameter P6.0.  
3. Press 
the 
ENTER key to select parameter 6.1.  
4. Press 
the 
MI/
S or MR/T keys to select the cell 
constant between K = 1.0, 0.1, or 10.  
5. Press 
the 
ENTER key to confirm selection and to 
return to the subgroup menu. Press the CAL/MEAS 
key to return to measurement mode.  
 
 
 
 
NOTES 
When using a cell of K = 0.1, the lowest measuring range 
will be: 0 to 1.999 
µS (0 to 0.999 ppm).  The 0 to 199.9 ms 
(10 to 200 ppt) range will not be accessible.  
 When using a cell of K = 10, the highest measuring range 
will be: 0 to 1999 mS (100 to 2000 ppt).  The 0 to 19.99 
µS 
(0.00 to 9.99 ppm) range will not be accessible.  
Figure 32 - Selecting cell constant