eutech-instruments waterproof cyberscan pc 10 phconductivity meter User Manual

Page of 31
Instruction Manual 
PC 10 
- 13 - 
4.6. 
Calibrating for Conductivity: 
1.  Pour out two separate portions of your calibration 
standard and one of DI water into separate clean 
containers.  
2.  If necessary, press MODE key to select 
Conductivity Mode. The 
µS and mS indicator will 
appear on the right side of the display.  
3.  Rinse your probe with DI water, then rinse the 
probe in one portion of the calibration standard. 
Immerse the probe into the second portion of the 
calibration standard. The meter auto ranging 
function selects the appropriate conductivity range 
(four ranges are possible). Be sure the probe is 
free from air bubbles and sufficient calibration 
standard is used to cover the conductivity sensor. 
Wait for the reading to stabilize. The READY 
indicator lights when the reading is stable. 
See Figure 18 
4.  Press the CAL/MEAS key. The CAL indicator 
appears above the primary display. The primary 
display shows the measured reading and the 
secondary display shows the temperature.  
See Figure 19. 
5. Press 
the 
▼ or ▲ key to scroll to the value of your 
conductivity standard. Press and hold the ▼ or ▲ 
keys to scroll faster. See Figure 20. 
MEAS
READY
°C
ATC
mS
CAL
READY
°C
ATC
mS
CAL
READY
°C
ATC
mS
Figure 16: CAL Indicator
 
Figure 15: Stabilized measurement
 
Figure 17:  Adjust to standard value