5 .   M E A S U R E M E N T  
The READY indicator appears on the display when 
the readings stabilise.  It will turn off if the readings 
start to fluctuate.  
NOTE:  Be sure to remove the protective cap of the 
pH electrode before measurement.  
To take readings: 
1.  Rinse the probe with DI water before use to 
remove any impurities adhering to the probe 
body.  If the pH electrode has dehydrated, 
soak it for 30 minutes in tap water.   
2.  Switch on the meter.  The MEAS annunciator 
appears on the top center of the LCD.  The 
ATC indicator appears in the lower right hand 
corner to indicate Automatic Temperature 
Compensation.  See Figure 22.  
3. 
Dip the probe into the sample.   
NOTE:  When measuring a sample, make sure that 
the probe is immersed beyond the electrode band. 
See figure in Section 4.2.  Stir the sample using the 
probe gently to create a homogenous sample. 
Allow 
time for the readings to stabilize.  Note the readings 
on the display.  When the readings are stable, the 
READY annunciator appears.  
4.  To toggle between pH and conductivity 
readings, press the MODE key.  See Figures 
23 and 24.  
NOTE:  Conductivity readings are auto 
ranging and will automatically move to 
the correct range (four ranges 
possible).  
MEAS
pH
°C
ATC
Figure 19: Measurement mode
 
Figure 20: READY
 
Figure 21: Conductivity mode