eutech-instruments waterproof cyberscan pc 10 phconductivity meter User Manual

Page of 31
Instruction Manual 
PC 10 
- 3 - 
2.2. Keypad 
The large membrane keypad makes the instrument easy to use. Each button, when pressed, 
has a corresponding graphic indicator on the LCD.  See Figure 3.  
Key Function 
ON/OFF 
Powers on and shuts off the meter.  The meter powers on in the mode it was  
shut off in.  For example, if you shut the meter off in Conductivity measurement 
mode, the meter will be in Conductivity measurement mode when you switch 
the meter on.  
HOLD 
Freezes the measured reading.  To activate, press HOLD while in 
measurement mode.  To release, press HOLD again.  
MODE 
Selects the measurement parameter (conductivity or pH).  Press MODE to 
toggle between pH and Conductivity mode.  
CAL/MEAS 
Toggles user between Calibration and Measurement mode.  
If you were in Conductivity Measurement mode, press CAL/MEAS to enter 
Conductivity Calibration mode.  
If you were in pH Measurement mode, press CAL/MEAS to enter pH 
Calibration mode.  
NOTE:  Temperature calibration is accessible from pH Calibration mode; see 
pages 10 for instructions.  
ENTER 
Press to confirm your calibration values in Calibration mode.  
 /   
Scrolls values up and down in Calibration mode.  
In pH mode,   /   scrolls through the auto buffer values 4.00, 7.00 and 
10.00.  
In Conductivity and Temperature Calibration mode,   /   lets you increase 
or decrease the calibration value.  
 
 
Figure 3: Keypad membranes of Waterproof PC 10