TROUBLESHOOTING HINTS 
 
Symptom                 Possible Causes          
 
Next Step 
Out of Range            defective meter           
 
check meter with shorting strap 
Reading                                    
 
 
 
(see meter instruction manual) 
 
                         
defective electrode           
 
check electrode operation 
 
                         
electrodes not          
 
 
unplug electrodes and reset  
                          
plugged in properly    
 
                         
reference electrode       
 
be sure reference electrode 
                         
not filled              
 
 
is filled 
 
                         
air bubble on membrane          
 
remove  bubble  by  redipping 
electrode 
 
                        
electrodes not          
 
 
put electrodes in solution 
                         
in solution           
 
 
Noisy or Unstable 
defective meter           
 
check meter with shorting strap 
Readings (readings                            
continuously or 
air bubble on membrane 
 
remove bubble by re-dipping electrode 
rapidly changing)                 
 
 
       
electrode exposed        
 
soak electrode in iodide standard 
                         
to interferences          
 
 
 
                         
defective electrode       
 
replace electrode 
 
                         
ISA not used              
 
use recommended ISA 
 
                         
meter or stirrer        
 
 
ground meter or stirrer 
                         
not grounded            
 
 
Drift (reading         
samples and standards   
 
allow solutions to come to room  
slowly changing         at different temperatures          
temperature before measurement 
in one direction)    
 
                 
electrode exposed        
 
check section entitled 
                         
to complexing agents                   
Complexation
 
                                              
                         
incorrect reference       
 
use recommended 
                         
filling solution        
 
 
filling solution 
 
Low Slope or   
standards contaminated     
 
prepare fresh standards          
No Slope               
or incorrectly made           
                      
                        
ISA not used              
 
use recommended ISA