eutech-instruments iodide epoxy User Manual

Page of 17
Instruction Manual 
 
Iodide Electrode
 
 
16
                         
standard used as ISA    
 
use ISA 
 
                         
electrode exposed        
 
check section entitled 
                         
to complexing agents               
Complexation
  
 
                         
air bubble on membrane          
 
remove  bubble  by  re-dipping 
probe 
                                      
 
"Incorrect Answer"    incorrect scaling 
 
 
plot millivolts on the linear axis. 
(but calibration       
of semi-log paper         
 
On the log axis, be sure that  
curve is good)                                                  
 
concentration numbers within 
 
        each 
decade 
are 
increasing 
with 
                                                    
 
 
 
increasing concentration. 
 
                         
incorrect sign          
 
 
be sure to note sign of the 
 
        millivolt 
number 
correctly 
 
                         
incorrect standards  
 
 
prepare fresh standards 
 
                         
wrong units used          
 
apply correct conversion factor: 
                                                
 
 
 
10
-3
M = 127.0 ppm as I
-1
 
 
                         
complexing agents       
 
check section entitled 
                         
in sample                      
 
Complexation