Silver/Silver Sulfide Electrode 
 
Instruction Manual 
 
24
TROUBLESHOOTING HINTS 
 
Symptom               
 
Possible Causes  
 
 
Next Step 
Out of Range   
 
defective meter 
 
 
check meter with 
Reading            shorting 
strap 
(see 
meter 
 
 
 
          instruction 
manual) 
 
 
 
                        defective electrode    
             check electrode operation 
 
 
 
 
 
 
 
 
 
 
 
 
   electrodes 
not     unplug 
electrodes 
 
 
 
 
plugged in properly   
 
and reset 
 
 
 
 
 
reference electrode 
 
 
be sure reference 
 
 
 
 
not filled 
 
 
             electrode is filled 
 
 
 
 
 
air bubble on  
 
 
remove bubble by 
 
 
 
 membrane 
  
 re-dipping 
electrode 
 
 
 
   electrodes 
not     put 
electrodes 
 
 
 
 in 
solution 
  
 in 
solution 
 
 
Noisy or Unstable 
 
defective meter 
 
 
check meter with 
Readings 
(readings 
         shorting 
strap 
continuously or  
 
rapidly changing)         
air bubble on  
 
 
remove bubble by 
 
 
 
 membrane 
  
 re-dipping 
electrode 
 
 
 
 
 
ISA not used  
 
 
use recommended ISA 
 
 
 
 
 
meter or stirrer 
 
 
ground meter or stirrer 
 
 
 
 
not grounded        
 
 
 
   defective 
electrode 
   replace 
electrode 
 
 
 
 
 
electrode exposed 
 
 
soak electrode in 
 
 
 
 
to interference 
 
 
silver or sulfide standards 
 
Drift (reading   
 
samples and standards 
 
allow solutions to 
slowly changing 
 
at different temperatures 
 
come to room temperature 
in 
one 
direction)          before 
measurement 
 
 
 
 
 
electrode exposed 
 
 
check section entitled 
 
 
 
 
to complexing agents  
 
Complexation 
 
 
 
   incorrect 
reference 
   use 
recommended 
 
 
   filling 
solution     filling 
solution