Instruction Manual 
 
Silver/Silver Sulfide Electrode 
 
25
Low Slope or   
 
standards contaminated 
 
prepare fresh standards 
No Slope 
 
 
or incorrectly made 
 
 
 
 
 
ISA not used  
 
 
use recommended ISA 
                                              
 
 
                        standard used as ISA   
 
use ISA 
 
 
 
                        electrode exposed 
 
 
check section entitled 
 
 
 
 
to complexing agents     
 
Complexation
  
 
 
 
 
 
air bubble on membrane 
 
remove bubble by 
                    
 
 
 
 
 
 
 
re-dipping probe 
 
 
"Incorrect Answer" 
 
incorrect scaling of    
 
plot millivolts on the linear 
(but calibration  
 
semi-log paper 
 
 
axis. On curve is good) 
 
 
 
 
 
 
 
 
 
the log axis, be sure that 
 
 
          concentration 
numbers 
 
 
          within 
each 
decade 
are 
 
 
          increasing 
with 
increasing 
 
 
          concentration 
 
 
 
 
 
incorrect sign  
 
 
be sure to note sign of 
 
 
          millivolt 
number 
correctly 
 
 
 
 
 
incorrect standards 
 
 
prepare fresh standards 
 
 
 
 
 
wrong units used 
 
 
apply correct conversion  
 
 
 
 
 
 
 
 
 
factor:  
 
 
          1.0X10
-3
M= 32.1 ppm S
-2
 
                                             
 
 
 
 
 
= 2.0X10
-3 
N (S
-2
); 
                                        
 
 
 
 
 
= 108 ppm Ag
+1
 
 
                                             
complexing agents 
 
 
check section entitled 
 
 
 
 in 
sample 
  
 
Complexation