4.1 
Simulating Liquid Parameter Check Result 
Simulating Liquid Dielectric Parameter Check Result
 
@ Muscle 2450 MHz 
Room Ambient Temperature =23
°C; Relative humidity = 35% 
Measured by: Ninous Davoudi 
f (MHz)
Temp. (°C) Depth (cm)
e"
Relative Permittivity (e'):
52.7
52.7762
0.14
± 5
14.9068
Conductivity (σ):
1.95
2.03175
4.19
± 5
Deviation (%)
Limit (%)
2450
21
15
Simulating Liquid
Parameters
Target
Measured
 
Liquid Check 
Ambient temperature: 23.0 deg. C; Liquid temperature: 21.0 deg C 
March 01, 2006 09:20 AM 
Frequency 
  e' 
  e'' 
2400000000. 
  52.9344 
  14.6795 
2410000000. 
  52.9127 
  14.7216 
2420000000. 
  52.8699 
  14.7759 
2430000000. 
  52.8501 
  14.8168 
2440000000. 
  52.8094 
  14.8432 
2450000000. 
  52.7762 
  14.9068 
2460000000. 
  52.7313 
  14.9246 
2470000000. 
  52.6793 
  14.9682 
2480000000. 
  52.6476 
  14.9936 
2490000000. 
  52.6200 
  15.0299 
2500000000. 
  52.5878 
  15.1083 
The
 c
onductivity (σ) can be given as: 
σ = ωε
e''
= 2 π f ε
e''
  
where  = target f * 10
ε
= 8.854 * 10
-12