REPORT NO: 06J10094-10 
DATE: March 2, 2006 
FCC ID: FDI-09102002-0
 
COMPLIANCE CERTIFICATION SERVICES 
Page: 12 of 39
This report shall not be reproduced except in full, without the written approval of CCS.
 
 
SYSTEM PERFORMANCE CHECK  
The system performance check is performed prior to any usage of the system in order to guarantee 
reproducible results.  The system performance check verifies that the system operates within its 
specifications of 
±10%.   
System Performance Check Measurement Conditions 
• 
The measurements were performed in the flat section of the SAM twin phantom filled with Body 
simulating liquid of the following parameters. 
•  The DASY4 system with an Isotropic E-Field Probe EX3DV3-SN: 3531 was used for the 
measurements. 
•  The dipole was mounted on the small tripod so that the dipole feed point was positioned below the 
center marking of the flat phantom section and the dipole was oriented parallel to the body axis (the 
long side of the phantom). The standard measuring distance was 10 mm (above 1 GHz) and 
15 mm (below 1 GHz) from dipole center to the simulating liquid surface. 
•  The coarse grid with a grid spacing of 15 mm was aligned with the dipole. 
For 5 GHz band - The coarse grid with a grid spacing of 10 mm was aligned with the dipole. 
•  Special 5 x 5 x 7 fine cube was chosen for cube integration(dx=dy=7.5mm; dz=5mm). 
For 5 GHz band - Special 8x8x8 fine cube was chosen for cube integration(dx=dy=4.3mm; dz=3mm) 
•  Distance between probe sensors and phantom surface was set to 4 mm.  
For 5 GHz band - Distance between probe sensors and phantom surface was set to 2.0mm 
•  The dipole input power (forward power) was 250 mW±3%. 
•  The results are normalized to 1 W input power. 
 
Reference SAR Values for body-tissue 
In the table below, the numerical reference SAR values of a SPEAG validation dipoles placed below the 
flat phantom filled with body-tissue simulating liquid are given.  The reference SAR values were 
calculated using the finite-difference time-domain method and the geometry parameters. 
 
Dipole Type 
Distance 
 (mm) 
Frequency 
(MHz) 
SAR (1g) 
[W/kg] 
SAR (10g) 
[W/kg] 
SAR (peak) 
[W/kg] 
D450V2 15  450  5.01 3.36 7.22 
D835V2 15  835  9.71 6.38 14.1 
D900V2 15  900  11.1 7.17 16.3 
D1450V2 
10  1450 29.6 16.6 49.8 
D1800V2 
10  1800 38.5 20.3 67.5 
D1900V2 
10  1900 39.8 20.8 69.6 
D2000V2 
10  2000 40.9 21.2 71.5 
D2450V2 10  2450  51.2 
23.7 97.6 
Note: All SAR values normalized to 1 W forward power.