Gossen Metrawatt METRAHIT CAL Calibrator, M244A Data Sheet
Product codes
M244A
METRAHIT
⏐
C
AL
Calibrator
3-349-440-03
2/7.08
GMC-I Gossen-Metrawatt GmbH
DQS Certified per
DIN EN ISO 9001 reg. no. 1262
Quality Management System
Calibration certificate included
System Components and Range of Applications
The METRAHIT CAL calibrator functions as a highly accurate
calibration and simulation instrument for electrical quantities. As a
handheld instrument, it’s suitable for precise, on-site calibration
and inspection tasks, as well as for test department and
laboratory work.
Thanks to its diverse functions, the highly flexible instrument can
be used in process engineering, control room and equipment
fabrication, general measuring technology and many other
applications. As a system component it can always be used in
calibration systems for calibrating measuring transducers, buffer
amplifiers, transmitters, temperature measuring and recording
instruments, controllers, signaling devices and indicators.
If a USB X-
calibration and simulation instrument for electrical quantities. As a
handheld instrument, it’s suitable for precise, on-site calibration
and inspection tasks, as well as for test department and
laboratory work.
Thanks to its diverse functions, the highly flexible instrument can
be used in process engineering, control room and equipment
fabrication, general measuring technology and many other
applications. As a system component it can always be used in
calibration systems for calibrating measuring transducers, buffer
amplifiers, transmitters, temperature measuring and recording
instruments, controllers, signaling devices and indicators.
If a USB X-
TRA
interface adapter is plugged on to the instrument
(accessory, see page 4), complete calibration procedures and
measuring-point oriented calibration cycles can be transferred
from a PC, saved to memory and accessed by simply pressing a
key. The calibrator setup procedure can thus be significantly
shortened, and erroneous settings can be avoided.
Optional METRAwin90-2 software simplifies programming,
controls data transfer to the calibrator, receives the measurement
data of any interconnected multimeter from a transducer output
and executes a comparison of the targeted and the actual
situation. Acquired values can be printed out as a calibration
report by the PC.
measuring-point oriented calibration cycles can be transferred
from a PC, saved to memory and accessed by simply pressing a
key. The calibrator setup procedure can thus be significantly
shortened, and erroneous settings can be avoided.
Optional METRAwin90-2 software simplifies programming,
controls data transfer to the calibrator, receives the measurement
data of any interconnected multimeter from a transducer output
and executes a comparison of the targeted and the actual
situation. Acquired values can be printed out as a calibration
report by the PC.
Universal Calibration Standard
Integrated electronics generate mV, V and mA signals. Beyond
this, they’re capable of simulating thermo-voltages for various
types of thermocouples for predefined temperatures (
this, they’re capable of simulating thermo-voltages for various
types of thermocouples for predefined temperatures (
°C or °F), as
well as resistance values for various Pt and Ni temperature
sensors.
sensors.
Frequency and Pulse Run Generator
Continuous frequency signals can be transmitted by the
METRAHIT CAL for testing SPCs, energy metering devices, flow
rates and more. Amplitude is adjustable for the generated
square-wave pulses, which are used to simulate sensor pulses.
Predefined pulse runs are also transmitted at a given frequency.
METRAHIT CAL for testing SPCs, energy metering devices, flow
rates and more. Amplitude is adjustable for the generated
square-wave pulses, which are used to simulate sensor pulses.
Predefined pulse runs are also transmitted at a given frequency.
Calibration and Simulation
Measuring transducers with a wide variety of input signals
(voltage, thermo-voltage, RTD and 2-wire resistance sensors etc.)
can be directly connected and calibrated. If a multimeter is used
(e.g. METRAHIT
(voltage, thermo-voltage, RTD and 2-wire resistance sensors etc.)
can be directly connected and calibrated. If a multimeter is used
(e.g. METRAHIT
⏐
X-
TRA
), the respective values can be measured at
the measuring transducer’s output, transmitted to a PC via an
adapter if desired, displayed with the help of optional
METRAwin90-2 software and compared with the appropriate
calibration specifications. Setpoint values and actual values are
displayed, or printed as a certificate. When operated in the “mA
sink” mode, the METRAHIT CAL simulates a 2-wire transmitter and
retrieves the selected current value from the measuring sequence.
adapter if desired, displayed with the help of optional
METRAwin90-2 software and compared with the appropriate
calibration specifications. Setpoint values and actual values are
displayed, or printed as a certificate. When operated in the “mA
sink” mode, the METRAHIT CAL simulates a 2-wire transmitter and
retrieves the selected current value from the measuring sequence.
•
Universal calibrator, simulator
mA / mV
mA / mV
… V / °C/°F (Pt100/1000, Ni100/1000,
thermocouples: J, L, T, U, K, E, S, R, B, N) / 5
… 2000 Ω
•
Frequency and pulse run generator: 1 Hz
… 1000 Hz
•
Ramp and staircase functions
•
Easy operation
•
Interface and optional METRAwin90-2 calibration software
•
Transmitter simulator (sink: 0
… 24 mA)
•
DKD calibration certificate included as standard feature
•
Rugged, EMC compliant design