10 Error messages
Display
Description
Remedy
Zero range exceeded
Unload the balance
A/D converter outside range
Unload the balance; check if
the weighing is correctly
placed and screwed on
the weighing is correctly
placed and screwed on
Displayed weight is
permanently changing
permanently changing
Avoid air draughts / air
movement as well as
vibrations of the table and of
the floor
movement as well as
vibrations of the table and of
the floor
Overload
Unload the balance and
adjust again
adjust again