11 Instant Help 
 
In case of an error in the program process, briefly turn off the balance and disconnect 
from power supply. The weighing process must then be restarted from the beginning. 
 
Fault 
Possible cause 
 
 
The displayed weight does not glow. 
  The balance is not switched on. 
 
  Batteries are inserted incorrectly or empty 
 
  No batteries inserted. 
The  displayed  weight  is  permanently 
changing 
  Draught/air movement 
 
  Table/floor vibrations 
 
  Weighing plate has contact with other 
objects. 
  Electromagnetic fields / static charging 
(choose different location/switch off 
interfering device if possible) 
 
 
The weighing value is obviously  
  The display of the balance is not at zero 
wrong 
  Adjustment is no longer correct. 
 
  Great fluctuations in temperature. 
  The balance is on an uneven surface. 
  Electromagnetic fields / static charging 
(choose different location/switch off 
interfering device if possible) 
 
Should other error messages occur, switch balance off and then on again. If the error 
message remains inform manufacturer.