Kern SFE 15K5IPMParcel scales Weight range bis 15 kg SFE 15K5IPM User Manual

Product codes
SFE 15K5IPM
Page of 13
Procedure 
 
Ö
  Confirm menu setting  
  by 
 .  
Ensure that there are no objects on the weighing plate.  
Ö
  Wait for stability display, then press 
.  
Ø
 
Ö
  The currently set adjustment weight will be displayed. 
Ö
  Either use the displayed adjustment weight or change it with  
 and 
, the active digit is flashing.  
Ö
 Confirm by 
, „LoAd“ will be shown. 
 
 
Ö
  Carefully place adjusting weight in the centre of the weighing 
plate. Wait for stability display, then press 
Ö
  After the adjustment the balance will carry out a self-test. 
Remove adjusting weight during selftest, balance will return 
into weighing mode automatically. 
An adjusting error or incorrect adjusting weight will be indi-
cated by the error message; repeat adjustment procedure. 
2  Linearization 
Linearity shows the greatest deviation of a weight display on the scale to the value of 
the respective test weight according to plus and minus over the entire weighing 
range. 
If linearity deviation is discovered during a testing instrument control, you can im-
prove this by means of linearization. 
 
•  Carrying out linearization is restricted to specialist staff possessing well 
acquainted with the workings of weighing scales. 
•  The linearisation is locked for verified weighing systems.  
In order to unlock the access, the seal must be destroyed and the 
jumper on the printed circuit board must be fitted. 
Attention: 
After destruction of the seal the weighing system must be re-verified by 
an authorised agency and a new verification wire/seal mark fitted before 
it can be reused for applications subject to verification. 
•  The test weights to be used must be adapted to the weighing scale’s 
specifications. 
•  Observe stable environmental conditions. Stabilisation requires a certain 
warm-up time. 
•  After successful linearization you will have to carry out calibration; see 
chapter “testing instruments control” 
8  
 