Kern SFE 15K5IPMParcel scales Weight range bis 15 kg SFE 15K5IPM User Manual

Product codes
SFE 15K5IPM
Page of 13
 
Procedure: 
Ö
  Call-up menu setting  
, see chap. 1. 
Ö
 Acknowledge with 
Ensure that there are no objects on the weighing plate. 
Ø
 
Ö
  Wait for stability display „STABLE“, then press 
. When 
“LoAd 1“ is displayed, put the first adjustment weight (1/3 
max) carefully in the centre of the weighing platform. 
Ö
  Wait for stability display „STABLE“, then press 
. When 
“LoAd 2“ is displayed, put the second adjustment weight (2/3 
max) carefully in the centre of the weighing platform. 
Ö
  Wait for stability display „STABLE“, then press 
. When 
“LoAd 3“ is displayed, put the third adjustment weight (1/3 
max) carefully in the centre of the weighing platform. 
Ö
  Wait for stability display „STABLE“, then press 
Ö
  After the adjustment the balance will carry out a self-test. 
Remove adjusting weight during selftest, the appliance will 
return into weighing mode automatically.  
 
An adjusting error or incorrect adjusting weight will be indi-
cated by the error message; repeat adjustment procedure. 
 
KFE-TM-ZB-def-1110