6.4  Initial Commissioning 
In order to obtain exact results with the electronic balances, your balance must have 
reached the operating temperature (see warming up time chap. 1). 
The weighing scale must be connected to the power supply (batteries) during this 
warm-up time. 
The accuracy of the balance depends on the local acceleration of gravity.  
Strictly observe hints in chapter Adjustment. 
 
6.5  Protection type IP65 
Designed for temporary contact with liquids. Use a damp cloth for cleaning. 
Dustproof. 
 
7  Adjustment 
 
As the acceleration value due to gravity is not the same at every location on earth, 
each balance must be coordinated -  in compliance with the underlying physical 
weighing principle - to the existing acceleration due to gravity at its place of location 
(only if the balance has not already been adjusted to the location in the factory).  
This adjustment process must be carried out for the first commissioning, after each 
change of location as well as in case of fluctuating environment temperature. To 
receive accurate measuring values it is also recommended to adjust the balance 
periodically in weighing operation. 
 
7.1  Adjustment 
Carry out adjustment as near as possible to the maximum load of the scale (See 
chpt. 1 “Techn. data”). The accuracy of the weight used for adjustment has to 
correspond approximately to readability d of the scale, or rather closer than that. 
Info about test weights can be found on the Internet at: http://www.kern-sohn.com 
 
Procedure when adjusting: 
Observe stable environmental conditions. A warming up time (see chapter 1) is 
required for stabilization.