Kern FFN 3K0.5IPNParcel scales Weight range bis 3 kg FFN 3K0.5IPN User Manual

Product codes
FFN 3K0.5IPN
Page of 31
19 
 
FFN-N-BA-e-1222 
 
7.1.1  Adjusting non-verifiable models 
 
Display 
Operation 
 
 
 
  Start balance by pressing 
 
  Whilst the weighing scale is carrying out a self-test,  
(00…- 99…) press 
 
and hold until the message “F1 
CAL“ appears. 
 
 
 
   Press 
in the display appears „UnLod“. 
 
 
 
(Example) 
   Press 
 again 
the weight value will be displayed. 
 
 
 
 
 
 
 
  Enter value for adjustment weight (See chpt 1) by 
 
  Position weight 
  Press 
 
Remove weight whilst scale is carrying out self-test 
Weighing scale changes to zero display. 
This completes the adjustment process.