FFN-N-BA-e-1222 
7.2  Verification 
 
General introduction
According to EU directive 90/384/EEC or 2009/23EG balances must be officially 
verified if they are used as follows (legally controlled area): 
a)  For commercial transactions if the price of goods is determined by weighing. 
b)  For the production of medicines in  pharmacies  as well as for analyses in  the 
medical and pharmaceutical laboratory. 
c)  For official purposes 
d)  For manufacturing final packages 
In cases of doubt, please contact your local trade in standard. 
 
Verification notes
An  EU  type  approval  exists  for  balances  described  in  their  technical  data  as 
verifiable.  If a balance is used where obligation to verify exists as described above, it 
must be verified and re-verified at regular intervals. 
Re-verification of a balance is carried out according to the respective national 
regulations. The validity for verification of balances in Germany is e.g. 2 years.  
The legal regulation of the country where the balance is used must be observed! 
 
 
Verification of the balance is invalid without the seal. 
The seal marks attached on verified balances point out that the balance may 
only be opened and serviced by trained and authorised specialist staff. If the 
seal mark is destroyed, verification looses its validity. Please observe all 
national laws and legal regulations. In Germany a re-verification will be 
necessary.