6.2.3  Units 
 
Different foreign weighing units are integrated into the various balance models .  
Change the Units with the  
 key. 
Press an hold the 
 key , few seconds later change the unit. 
 
 
Display 
 
Conversion factor 
1 g = 
Gramm  
1. 
Unze oz 
0.035273962 
Troy Unze 
ozt 
0.032150747 
Pennyweight dwt  0.643014931 
 
6.2.4  Dosage und Zero-tracking 
When the Auto-Zero-function is activated, any slight changes of the zero readout are 
automatically tared. 
In the event that small quantities are removed or added to the material to be 
weighed, incorrect weighing results can be displayed due to the “stability compensa-
tion“ in the balance. (Example: Slowly draining fluids from a container on the bal-
ance.)   
Having dosages with slight changes of weight it is recommended to deactivate this 
function. 
 
With switched off Zero-Tracking the display however is more unstably. 
 
Activate/deactivate Zero-Tracking 
Ö  When the balance is switched off, press the OFF -key and keep it pressed 
Ö  Do not release the OFF -key. Press also the ON/TARE-key and keep it pressed  
Ö Keep the ON/TARE-key pressed, however release the OFF -key 
Ö Press OFF -key anew 
Ö  Keep both keys pressed until in the display “tr” appears 
Ö  Release both keys. The balance is situated in the menu. 
Ö Press the ON/TARE-key until „tr“ appears  
Ö  Confirm with the OFF key  