SOEKS IMPULSE Magnetic field-analysis apparatus, Magnetic field tester 9 Data Sheet

Product codes
9
Page of 52
 EMF in LIVE AREA
 
Set the "EMF in LIVE AREA" mode 
 
Bring the instrument into the monitored area, holding it with your 
fingers at arm's length at a height of 1–2 meters above the ground or other 
surface. Keep instrument in the monitored area not less than 10 seconds prior 
to reading out the result. Measures should be taken in case of appearance of 
"HIGH LEVEL  OF  ELECTRIC  FIELD"  or  "HIGH LEVEL  OF 
MAGNETIC FIELD" messages on the red background. 
 
The possible reasons for threshold crossing are: proximity of 
power cable, wiring defect, proximity to the electric devices (e. g. elevator 
motors), proximity to the base stations, crossing of the tracks of radio relay 
stations, proximity to the power lines, etc. Long stay of people and domestic 
animals in such areas should be avoided; in case of significant excess staying 
there may result in electrical shock.
 EMF of PC
 
Set the "EMF of PC" mode 
 
Take measurement according to the procedure specified in 
Appendix 3.
 
Measurement time in this case should be not less than 30 seconds. 
For  the  purpose  of  reducing  electromagnetic  field  distortion  generated  by 
human, it is advisable to fix the instrument on an electrically neutral stand, e. 
g. plastic chair, for the time of measurement.
 Localization of hidden electric wiring
 
The instrument can be used for the localization of a hidden 
electric wiring in walls or furniture.
 
To do this, it is convenient to use graphic electric and magnetic 
field level indicators, displaying not average but instant value of the 
corresponding field parameter.
 
Select  the  required  range  for  instant  field  level  display  by  an 
experimentation. If the field is weak, use «EMF of PC» mode; if the signal 
exceeds the display range, use other modes.
 
 
 
 
-41-