Siemens Sitrans F FX300 User Manual

Page of 80
8
 TECHNICAL DATA 
68 
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02   A5E02100423
Operating conditions
Temperature
Temperature
Temperature
Temperature
Product
-40…+240°C / -40…+465°F
Ambient
Non-Ex: -40…+85°C / -40…+185°F
Ex: -40…+60°C / -40…+140°F
Storage
-50…+85°C / -58…+185°F
Pressure
Pressure
Pressure
Pressure
Product
Max. 100 bar / 1450 psi; Information on higher pressures on request.
Ambient
Atmosphere
Chemical properties
Chemical properties
Chemical properties
Chemical properties
Density
Taken into consideration when sizing.
Viscosity
< 10 cP
Reynold's number
10000...2300000
Recommended flow velocities
Recommended flow velocities
Recommended flow velocities
Recommended flow velocities
Liquids
0.3…7 m/s  /  0.98…23 ft/s (optional up to 10 m/s / 32.8 ft/s taking 
cavitation into account)
Gases and steams
2.0…80 m/s / 6.6…262.5 ft/s
DN15: 3.0…45 m/s / 9.8…148 ft/s; DN25: 2.0…70 m/s / 6.6…230 ft/s
For detailed information, see chapter "Flow tables".
Other conditions
Other conditions
Other conditions
Other conditions
Protection category
IP 66/67
Installation conditions
Inlet run
≥ 20 x DN (without disturbing flow, after pipe narrowing, after a single 
90° bend)
≥ 30 x DN (after a double bend 2x90°)
≥ 40 x DN (after a double three-dimensional bend 2x90°)
≥ 50 x DN (after control valves)
≥ 2 DN before flow straightener; ≥ 8 DN after flow straightener 
(specified values apply only to original ≥ 20 DN inlet run)
Outlet run
≥ 5 x DN
Dimensions and weights
For detailed information refer to chapter "Dimensions and weights".
Materials
Measuring sensor and process 
connections
Standard: 1.4404/316L
Option: Hastelloy
®
 C-22 on request
Converter housing
Die-cast Aluminium
Pressure sensor gasket
Standard: FPM
Option: FFKM
Measuring tube gasket
Standard: 1.4435/316L
Option: Hastelloy
®
 C-276
Selection depends on measuring sensor material/medium.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book  Page 68  Monday, June 22, 2009  8:17 AM